OMS LaserPoint LP01 Laser Vibrometer Specifications
The OMS LaserPoint LP01 Vibrometer is based on a patented electro-optical configuration. The system consists of two components, an optical head and an electronic controller, which are connected by a cable.
The optical head contains a near-infrared laser diode with a built-in high speed photo-detector, an acousto-optic modulator, and a red laser diode to aid in aiming the beam. The controller contains a demodulator, a signal strength indicator, two selectable velocity ranges, and an array of low pass filter options.
- Velocity Range
- 5 μm/s to 800 mm/s 1
- Vibration Frequency Range
- 0.01 Hz to 20 kHz 2
- Working Distance
- 1 cm to 5 m
- Displacement Range
- 0.04 nm (at 20 kHz) to 120 mm (at 1 Hz)
- Optics
- Collimated (No Focusing Needed)
- Surface Reflectivity
- Realistic Surfaces
- Signal Output
- Analog Velocity and 10.7 MHz FM Outputs
- Dimensions
- Laser Head: 24 x 11.4 x 7.6 cm
- Electronic Controller: 30 x 22 x 6 cm
- Weight
- Laser Head: 1.4 kg
- Electronic Controller: 3.9 kg
- Low Pass Filters
- 1,2,5,10,20 kHz
- Output Voltage (max)
- ± 10 Volts
- Laser
- 780 nm, <15mW, Class IIIb
- 650 nm, <1mW, Class II
- Power Requirement
- 100 to 230 VAC at 50/60 Hz
- Temperature Range
- 3 - 45°C
1 Higher velocities are available with the LaserPoint. Please contact us for more information.
2 Frequencies above 20 kHz can be measured with the LaserPoint. Please contact us for more information